Olympus 38DLP-X-U-ET-E-EN

Olympus 38DLP-X-U-ET-E-EN

GAUGE ULTRASONIC THKNS 0.003 TO 25 IN

Price: loading
Product Information:

Additional Specifications

Olympus 38DLP-X-U-ET-E-EN

Olympus Ultrasonic Thickness Gauge (Without Probe), Series: 38DL, 0.003 to 25 in Measuring, 0.0001 in Resolution, Liquid Crystal Display, 2 to 30 MHz Frequency, 0.508 to 13.998 mm/us Velocity

Measuring Range      :      0.003 to 25 in
Series      :      38DL
Frequency Rating      :      2 to 30 MHz
Item Name      :      Ultrasonic Thickness Gauge (Without Probe)
Resolution      :      0.0001 in
Display      :      Liquid Crystal
Velocity      :      0.508 to 13.998 mm/us

The 38DL PLUS is an innovative instrument that signals a new era in ultrasonic thickness gaging. Ideally suited for almost every ultrasonic thickness application, this handheld thickness gage is fully compatible with a full line of dual and single element transducers. The versatile 38DL PLUS can be used in applications ranging from wall thinning measurements of internally corroded pipes with dual element probes to very precise thickness measurements of thin or multilayer materials with single element transducers. The 38DL PLUS comes standard with many powerful but easy-to-use measurement features and a host of application-specific software options. Its sealed case is designed to meet IP67 requirements to withstand the rigors of very wet or dusty environments. The color transflective VGA display provides superior readability from bright sunlight to complete darkness. It features a simple, ergonomic keypad that can be operated with the left or right hand for easy access to all functions.

  • Dual and single element transducer compatibility
  • Corrosion thickness gaging with dual element transducers
  • THRU-COAT® and Echo-to-Echo measurements on painted and coated surfaces
  • Standard resolution of 0.01 mm or 0.001 in for all transducers
  • High resolution software option of 0.001 mm or 0.0001 in with single element transducers 2.25 to 30 MHz
  • Thickness, velocity, and time-of-flight measurements
  • Differential mode and reduction rate mode
  • Time-based B-scan mode, 10,000 reviewable readings per scan
  • Olympus High Dynamic Gain technology with digital filters
  • V-Path builder for custom V-path compensation
  • Designed for EN15317